Week by week and bias against grade
Overview
Week by week and Bias against grade look at a whole laboratory, where the other Accuracy charts look at one CRM. Both compare each CRM result with its certified value, so they use only CRMs that have a certified value and a certified SD. They open from the Lab events card on the Overview, and they are in the Chart list on the Charts tab.
Where the data comes from
On the Overview they cover every date, every analysis suite and every analyte at the laboratory, whichever analytes the Overview assesses. On the Charts tab they follow the Filters: the Laboratory, the Analysis Suite and the Date range or despatches. The CRM type and CRM filters are not used, because both charts take every CRM. Week by week takes every analyte with the same method as the Analyte chosen. Bias against grade takes the Analyte chosen.
Week by week
Week by week has one row for each analyte and one square for each week. Each square is the median, over one week’s CRM results for one analyte, of (result − certified value) / certified SD.
- Weeks run Monday to Sunday by the date chosen under Date to use for data extraction: the laboratory return date unless you choose the send date. A square needs 3 results to be coloured.
- The five colours are 1 SD or more below, 0.5 to 1 SD below, within 0.5 SD, 0.5 to 1 SD above, and 1 SD or more above.
- Results whose CRM has no certified value or SD are left out, and the window says how many.
- An analyte’s method is the middle of its name, Element_Method_Units. The window opens on the method with the most results.
It shows how long a shift lasted and how many analytes it touched. Lab events, once looked for, are drawn on it as numbered lines.
Bias against grade
Bias against grade says what kind of bias a laboratory has. For one analyte and one period, each CRM gives one bias: its median result minus its certified value. Two shapes are fitted to those biases by least squares, each CRM’s misfit measured in its own certified SDs. A CRM result at or above the upper detection limit is not a measurement and is left out, as in the CRM checks.
- The same percentage at every grade: bias = k × certified value.
- The same amount at every grade: bias = c.
- Each shape’s misfit is the root mean square of what it leaves over, in certified SDs. The shape whose misfit is 0.6 of the other’s, or less, names the pattern. Otherwise the window says the bias has no clear shape.
- A CRM needs 5 results in the period. An analyte needs 4 such CRMs, the highest certified value at least 10 times the lowest.
- A pattern is looked for only when more than half the CRMs read 0.5 SD or more to the same side of certified. Otherwise the window says there is no consistent bias.
On the Overview the periods are every date and, once lab events have been looked for, the stretches between them. A stretch runs from the last day of one event to the day before the next begins. On the Charts tab the period is the Date range or the despatches chosen.
| Shape | What it means | What to ask the laboratory about |
|---|---|---|
| The same percentage at every grade | The bias scales with the result | Calibration, dilution and sample weight |
| The same amount at every grade | The bias is added to every result, so its percentage is largest at low grade | Its blank or background correction, contamination and the calibration’s zero |
The wording is always something to ask the laboratory. It is never a verdict.
At your cut-off
With a cut-off entered in Rules, under Grades you model, Bias against grade also gives the bias at that grade.
- With the same percentage at every grade, the bias at the cut-off is that percentage of the cut-off.
- With the same amount at every grade, it is that amount, and its percentage is the amount divided by the cut-off.
- Without a shape, it is the median percentage bias of the CRMs certified between half and twice the cut-off. With none there, it is the average of the two CRMs either side of the cut-off.
- A cut-off with no CRM near it and none on both sides is not judged. When CRMs are certified near it but have fewer than 5 results in the period, or no certified SD, the window names them and says what to do.
- A bias at the cut-off of 3% or more is a caution, and of 10% or more is not trusted yet. Both are set in Rules. The bias is judged as it is shown, to one decimal.
The same is worked out at the ore grade when one is entered, and judged by the same rules. Each period lists the despatches with QC results of the analyte returned in it, from the standards and the duplicates.
Precision at the cut-off
The duplicate pairs near the cut-off are those whose mean is between half and twice it. They must also be usable as the Precision check counts them: a mean of at least 10 times the detection limit.
- They are grouped by check stage, in the period chosen.
- The median RD% of the pairs, times the cut-off, says how far apart two results of the same material sit at that grade.
- A stage passes when at least 90% of its pairs are within its RD limit, warns from 80% and fails below. These are the Precision check’s rules, set in Rules.
- A stage with fewer than 10 pairs near the cut-off is too few to judge.
- The pairs whose two results fall on opposite sides of the cut-off are counted for information only. A sample close to the cut-off can change sides even when the laboratory repeats well.
Blanks against the cut-off
- The blanks of the analyte at the laboratory, in the period, are taken as reported. Blanks below detection count at half the detection limit.
- Their median and largest result are given as a percentage of the cut-off. The median includes what the blank material itself holds, so it is not judged.
- A blank 10% of the cut-off or more above the median points to contamination in its despatch. The 10% is the not-trusted share set in Rules.
Despatches not checked
- The despatches are those with any QC result at the laboratory in the period, from the standards and the duplicates.
- For each modelled analyte, the app counts those with no CRM, no blank and no duplicate of it, and names those with none of the three.
- A despatch with no QC sample at all is not in the QC data.
The Grades card
- Each despatch with QC at the laboratory is placed in the period of its first QC return date. A despatch returned during a lab event, between two periods, joins the period after it.
- For each modelled analyte it takes the outcome of the bias at the cut-off in that period, and at the ore grade when one is entered: the worse of the two. No CRM certified near the cut-off, and none on both sides, leaves it Not judged, unless the ore grade is a caution or worse. An ore grade that can’t be judged is said.
- Precision at the cut-off in that period can make it worse. A warning at a laboratory check stage is a caution, and a fail is Not trusted yet. Too few pairs near the cut-off changes nothing.
- Field duplicate stages are noted but don’t change the outcome. A stage is a field stage when it is typed as one on the Stages list, is a standard field stage (Sampling, Sampling Dust, Field Prep, Field Prep Duplicate Fraction, Lab Submission), or its name has field, sampl or dust in it. A field duplicate’s spread includes the sampling and splitting, so it doesn’t judge the laboratory.
- A despatch with no CRM, blank or laboratory duplicate of the analyte is Not trusted yet. Field duplicates alone don’t check the laboratory.
- The outcome is the worst of these, in the order Trusted, Trusted with a caution, Not judged, Not trusted yet. Every cause that isn’t Trusted is named, the worst first.
- A blank reading 10% of the cut-off or more above the usual level, and a despatch with no CRM of its own, are noted but don’t change the outcome. No blank limit is set in the Rules.
- The card counts each analyte’s despatches by outcome. An analyte’s colour, and the card’s, is the outcome most of its despatches have, a tie going to the worse. The export has one row per despatch, with each analyte’s outcome and why.
- The cut-offs are those entered in Rules for any analyte in the data, whether or not it is ticked in the Overview’s Analytes list.